RadiCS
Chapter 4 Acceptance Test / Constancy Test
41
English
Grayscale Check
Checks grayscale.
[Procedure]
1. Clic
k [Task] - [Grayscale check].
2. “Input tester name” will appear. Input tester’s name and click [OK].
Note
Input tester’s name in 31 characters.
3. Click [Proceed].
Tips
Click [Skip] to skip a monitor grayscale check onto the next monitor.
4. Follow the directions of the grayscale check.
5. Click [Next Monitor] to check grayscale of some other monitor. Click [Exit] to exit
the check.
Tips
For GS521-ST, if Grayscale Check has not been executed, a black image is displayed in full screen.
Uniformity Check
Checks white luminance uniformity of a whole screen.
[Procedure]
1. Clic
k [Task] - [Uniformity check].
2. “Input tester name” will appear. Input tester’s name and click [OK].
Note
Input tester’s name in 31 characters.
3. Click [Proceed].
Tips
Click [Skip] to skip a monitor uniformity check onto the next monitor.
4. Follow the directions of the uniformity check.
5. Click [Next monitor] to check uniformity of some other monitor. Click [Exit] to exit
the check.
Tips
Set up the schedule to test constancy regularly. See “Chapter 6. Scheduling” for details.
For GS521-ST, if Uniformity Check has not been executed, a black image is displayed in full screen.